Customization: | Available |
---|---|
Certification: | ISO9001, CCC |
Protection: | IP20 |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
GF302 Portable Multifunction Instrument Calibrator is suitable for power plant and power grid companies for the following function: measuring and testing department and instrumentation classes, national levels measuring and testing institutions, railway, petroleum, chemical industry and other large industrial and mining enterprises, scientific research units, etc. The core technology function with digital signal processor (DSP) and 16 high-speed digital converters composed of high precision work frequency communication terminal. The signal source is DSP and 16 high-speed digital-to-analog converters, it can control the sine wave and distortion wave signal source.
Electrical parameters | |
---|---|
Accuracy class | 0.05%, 0.1% |
Power supply | Single phase AC 220V±10% or 110V±10%, 50/60Hz |
Communication port | USB, RS232, RS485, LAN |
AC Voltage output | |
Range(U1,U2,U3) | 50V, 100V, 200V, 400V, 600V |
Adjustment range | (0 - 120)% RG |
Adjustment resolution | 0.01% RG, 0.1% RG, 1% RG, 10% RG |
Stability | 0.01% /1min |
Distortion | ≤0.2% (non-capacitive load) |
Max. output load | 25VA for each phase |
Accuracy | 0.05% RG |
AC Current output | |
Range(I1,I2,I3) | 0.5A, 1A, 2.5A, 5A, 10A, 20A |
Adjustment range | (0 - 120)% RG |
Adjustment resolution | 0.01% RG, 0.1% RG, 1% RG, 10% RG |
Stability | 0.01% /1min |
Distortion | ≤0.2% (non-capacitive load) |
Max. output load | 25VA for each phase |
Accuracy | 0.05% RG |
AC Power output | |
Active output stability | 0.01%RG/1min |
Reactive output stability | 0.02%RG/1min |
Active accuracy | 0.05% RG |
Reactive accuracy | 0.1% RG |
Frequency output | |
Adjustment range | 45-65Hz |
Adjustment resolution | 1Hz, 0.1Hz, 0.01Hz and 0.001Hz |
Resolution | 0.001Hz |
Accuracy | 0.002Hz |
Power factor output | |
Adjustment range | -1 to 0 to +1 |
Adjustment resolution | 0.0001 |
Resolution | 0.0005 |
Phase output | |
Adjustment range | 0°-359.999° |
Adjustment resolution | 10°, 1°, 0.1°, 0.01° |
Resolution | 0.001° |
Accuracy | 0.05° |
Harmonic configuration | |
Times | 2 to 31 |
Content | 0-40% |
Phase | 0°-359.999° |
Configuration error | (10% RD + 0.1%), RD refers to the configuration value of harmonic contents |
DC Voltage output | |
Range | 75mV, 75 V, 150 V, 300 V, 500V, 1000 V |
Adjustment range | (0-120)% RG |
Adjustment resolution | 0.01% RG, 0.1% RG, 1% RG, 10% RG |
Stability | 0.01% RG / 1 min |
Distorting | ≤0.2% (non-capacitive load) |
Output load | 25VA |
Accuracy | 0.05% RG |
Ripple contents | ≤1% |
DC Current output | |
Range | 0.5 A,1A, 2.5 A, 5 A, 10A, 20 A |
Adjustment range | (0-120)% RG |
Adjustment resolution | 0.01% RG, 0.1% RG, 1% RG, 10% RG |
Stability | 0.01% RG / 1min |
Distortion | ≤ 0.2% (non-capacitive load) |
Output load | 25VA |
Accuracy | 0.05% RG |
Ripple contents | ≤1% |
Energy Error | |
Active error | 0.05% RG |
Reactive error | 0.1% RG |
DC Input Voltage Measurement | |
Range | 0 to ±20V |
Measurement range | (0-120)% RG |
Accuracy | 0.01% RG |
Resolution | 0.001% RG |
DC Input Current Measurement | |
Range | 0-20mA |
Measurement range | (0-120)% RG |
Accuracy | 0.01% RG |
Resolution | 0.001% RG |
Mechanical parameters | |
Dimensions (W×H×D) (mm) | 460x430x185 |
Weight (kg) | 20 |
Environmental conditions | |
Working temperature | 0°C to 40°C |
Relative humidity | ≤85% |