Customization: | Available |
---|---|
Warranty: | Three Years |
Application: | Electromagnetic Instrument |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
Electrical parameters | |
---|---|
Accuracy class | 0.02%, 0.05% |
Power supply | AC 220 V±10% or 110 V±10%, 50/60Hz |
Power consumption | <30VA |
Warming-up time | <30min |
Voltage measurement | |
U1, U2,U3 | 6.25V, 12.5V, 25V, 50V, 100V, 200V, 400V, 800V (switch automatically), max 1000V |
Range | (0-120%)RG |
Resolution | 0.01%RG |
Accuracy | 0.02%RD or 0.05%RD |
Current measurement | |
I1, I2, I3 | 0.25, 0.5, 1, 2.5, 5, 10, 25, 50, 100A(switch automatically) |
Range | (0-120%)RG |
Resolution | 0.01%RG |
Accuracy | 0.02%RD or 0.05%RD |
Power measurement | |
Active accuracy | 0.02%RD or 0.05%RD |
Reactive accuracy | 0.05%RD or 0.1%RD |
Apparent accuracy | 0.02%RD or 0.05%RD |
Energy measurement | |
Active accuracy | 0.02%RD or 0.05%RD |
Reactive accuracy | 0.05%RD or 0.1%RD |
Phase measurement | |
Range | 0.00°-359.99° |
Resolution | 0.001° |
Accuracy | 0.02° (voltage ≥50V and current ≥250mA) |
Frequency measurement | |
Range | 45-65Hz |
Resolution | 0. 001Hz |
Accuracy | 0.002Hz |
Power Factor measurement | |
Range | -1.0 ~ 0 ~ +1.0 |
Resolution | 0.0001 |
Accuracy | 0.0005 |
Electric energy pulse output | |
Pulse constant range | 1-2880000000 |
Output frequency of max. pulse | 600 kHz |
Load capacity | >2mA |
Default pulse output frequency | 10kHz-600kHz |
Energy pulse Input | |
Input range of pulse constant | 600-700000 |
Setting range of test pulse | 1-999999999 |
Max. pulse receiving frequency | 2MHz |
Pulse input level | 5V |
Voltage/current harmonics measurement | |
Times | 2-31 |
Error | 0.05% |
Distortion degree error | 0.05% |
Transducer measurement | |
DC voltage range | ±1, ±5, ±10, ±20 V |
Accuracy | 0.01% |
DC current range | ±1, ±2.5, ±5, ±10, ±20mA |
Accuracy | 0.01% |
Ripple error | 1% |
DC measurements (option) | |
DC voltage | 50V, 100V, 200V, 400V, 800V (switch automatically), max 1000V |
Range | (0 -120%) RG |
Resolution | 0.01% RG |
Accuracy | 0.02% RD |
DC current | 0.1A, 0.3A, 1A, 3A, 10A, 30 A(switch automatically) |
Range | (0-120%) RG |
Resolution | 0.01% RG |
Accuracy | 0.02% RD |
Mechanical parameters | |
Dimensions (W×H×D) (mm) | 440×360×160 |
Weight (kg) | 10 |
Environmental conditions | |
Operating temperature | 10°C to 30°C |
Relative humidity | ≤85% |