| Customization: | Available |
|---|---|
| After-sales Service: | 24 Hours Service |
| Warranty: | 3 Years |
| Shipping Cost: | Contact the supplier about freight and estimated delivery time. |
|---|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency

Multi secondary winding test optional;
Produce a word or PDF file test report;
With battery optional, working one week;
PT test, easy to test, all the tests result quickly;
Portability: weight 4.8kg, the best light pt analyzer
Tests such as ratio, phase, polarity, capacitive ratio;
With thermal printer, printing test results in the field;
Parameters such as knee point current and voltage;
Parameters such as 10% error curve, 5% error curve;
The data can be displayed and analyzed after transferred to PC;
Tests of various types of measurement and protection VTs and CVTs;
Using of advanced power technology, the test knee point reaches up to 80kV;
No external other auxiliary equipment, stand-alone to complete all test items;
Storage 10000 groups of test data which would not be lost if the device is power off;
Power plant;
Electrical laboratory;
Metrological service center;
Voltage transformer factory;
Power engineering commissioning company;
Electricity power bureau & power company;
National Metrology and testing department;
Electrical Department of industrial and mining enterprises;
| I. Voltage Transformer (PT) |
|---|
| 1. Excitation characteristic test |
| 2. Transformation ratio test |
| 3. Polarity |
| 4. Ratio error, phase error |
| 5. Degauss |
| 6. Calculation of knee point value |
| 7. burden test |
| 8. Resistance test |
| Electrical parameters | ||
|---|---|---|
| PT phase error measurement | Accuracy | ±1min (typical) / 3 min (guaranteed) |
| Resolution | 0.1min | |
| PT ratio error measurement | Range | 1-15000 |
| Accuracy | ≤0.05% | |
| LCD display | 7' inch TFT touch color LCD | |
| Cable Length | Primary 5m; Secondary 5m; others customized | |
| Communication port | USB, RS232, WIFI | |
| PC control software | Yes, Optional | |
| Printer | Yes, Thermal printer | |
| Standards | ||
| Reference standards | GB1207-2006, GB1208-2006, GB16847-1997 IEC60044-6, IEC61869-1,3, ANSI/IEEE C57.13 |
|
| Safety standards | GB 4793.1-2007 | |
| EMC | EMC standard 89/336/EEC FCC Subpart B of Part 15 Class A IEC 1000-4-2/3/4/6 |
|
| Mechanical parameters | ||
| Overall dimension (L x W x H) (mm) | 280 x 250 x 160 | |
| Weight (kg) | ≤4.8 | |
| Environmental conditions | ||
| Relative humidity | Relative humidity 5%-95% not condensing | |
| Operating temperature | -10°C to +50°C | |
| Storage temperature | -20°C to +70°C | |
| Altitude | ≤2000m; If the altitude is greater than 2500m, the instrument needs to be customized | |