Customization: | Available |
---|---|
After-sales Service: | 24h, 7days |
Warranty: | 3 Years |
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
GF6018A high accuracy portable ac/dc current source and voltage source is according to verification regulation JJG124-2005 "ammeter, voltmeter, power meter and resistance meter calibration regulations, JJF1075-2001" clamp ampere ammeter calibration the requirements of the "standards for the design of 0.05 magnitude AC and DC standard units.The source of the signals of the DSP and 16 high-speed digital-to-analog converters can control of the sine wave, distortion wave source.The display USES graphics standard 240 X 128 LCD monochrome screen
1.GF6018A high accuracy portable ac/dc current source and voltage source consists of a single-phase AC voltage standard resource, current standard resource and DC voltage, current standard resource of standard resource.
2.Have calibration function, which can according to the standard of external value, fixed standard resource output values.
3.The setting of the range for verification 1999 (9), 3999 (9) a display of the digital multimeter.
4.Manual and semi-auto inspection 0.2% and below levels of digital multimeter, an analogue meter installation type digital instrument, including DC analog output of digital instrument;All kinds of clamp ampere ammeter;Verification clamp ampere ammeter range for 200 A, 600 A, 1000 A.
5.To generate containing 2 ~ 31 harmonic wave distortion, the harmonic number, number, amplitude and phase of the harmonic wave to base wave all can be programmed control.
6.With RS232 interface and large capacity non-volatile memory, can store 500 blocks are inspection table test data.Through the PC software, the PC control this device automatic and manual inspection, and the results are processing and management.
Technical Data
Electrical parameters | ||
---|---|---|
Power supply | Single phase AC 220V±10% or 110V±10%, 50/60HzAC | |
AC Voltage output | ||
range | 200mV, 400mV, 2000mV, 4000mV, 20V, 100V, 200V, 500V, 1000V |
|
Range 20V-1000V | Accuracy | ±(0.03%RD+0.02%FS) |
Stability | ≤0.01% FS /60s | |
Range 200mV-4000mV | Accuracy | ±(0.3%RD+0.2%FS) |
Stability | ≤0.04% FS /60s | |
Load capacity | 20VA | |
Waveform distortion | ≤0.3% | |
Adjustable Range | 0-120% FS (1000V not included) | |
Fineness | 5×10-5 | |
AC Current Output | ||
Range | 2mA, 4mA, 20mA, 40mA, 0.2A, 0.5A, 2A, 5A, 10A, 20A | |
Range 0.2A-20A | Accuracy | ±(0.03%RD+0.02%FS) |
Stability | ≤0.01% FS /60s | |
Range 2mA-40mA | Accuracy | ±(0.3%RD+0.2%FS) |
Stability | ≤0.04% FS /60s | |
Load capacity | 20VA | |
Waveform distortion | ≤0.3% | |
Adjustable Range | 0-120% FS | |
Fineness | 5×10-5 | |
Frequency Output | ||
Range | 45-65Hz | |
Fineness | 0.001Hz | |
Accuracy | 0.01Hz | |
Harmonic Wave | ||
Times | 2-31 | |
Extent | 0-20% | |
Each harmonic phase fineness | 0.01·N (N is harmonic order) | |
DC voltage output | ||
Range | 200mV, 400mV, 2000mV, 4000mV, 20V, 40V, 100V, 200V, 400V, 1000V |
|
Load capacity | 20W | |
Adjustable range | 0-110% FS (1000V not included) | |
Adjustable fineness | 5×10-5 | |
Stability | ≤0.01% FS /60s (peak-peak value) | |
Accuracy | ±(0.03%RD+0.02%FS) | |
Ripple wave | ≤1% FS | |
DC current output | ||
Range | 20uA, 40uA, 200uA, 400uA, 2mA, 4mA, 20mA, 40mA, 0.2A, 1A, 2.5A, 5A, 10A, 20A |
|
Load capacity | 20W | |
Adjustable Range | 0-110% FS (1000V not included) | |
Adjustable Fineness | 5×10-5 | |
Stability | ≤0.01% FS /60s (peak-peak value) | |
Accuracy | ±(0.06%RD+0.04%FS) | |
Ripple wave | ≤1% FS | |
Resistance box | ||
Range | 0-200Ω, 200-400Ω, 0-2kΩ, 2kΩ-4kΩ, 0-20kΩ, 20kΩ-40kΩ, 0-200kΩ, 200kΩ-400kΩ, 0-2MΩ, 2MΩ-4MΩ |
|
Fineness | 0.02% | |
Range 0-2kΩ, 0-20kΩ, 0-200kΩ | ||
Range 0-2kΩ | Adjustable Range | 0-100%FS |
Accuracy | ±(0.12%RD+0.08%FS) | |
Range 0-200Ω,0-2MΩ | Adjustable Range | 0-100% FS |
Accuracy | ±(0.3%RD+0.2%FS) | |
Range 2kΩ-4kΩ, 20kΩ-40kΩ, 200kΩ-400kΩ | ||
Range 2kΩ-4kΩ | Adjustable Range | 5%-100%FS |
Accuracy | ±(0.12%RD+0.08%FS) | |
Range 200-400Ω, 2MΩ-4MΩ |
Adjustable Range | 5%-100% FS |
Accuracy | ±(0.3%RD+0.2%FS) | |
Clamp meter range | ||
200A | Accuracy | ±0.2% |
600A | Accuracy | ±0.5% |
1000A | Accuracy | ±1% |
Dot frequency square wave output | ||
Range | 1kHz, 5kHz, 10kHz, 50kHz, 100kHz, 200kHz, 1000kHz | |
Square wave amplitude | 4.5V | |
Setting accuracy | 5×10-5 | |
Mechanical parameters | ||
Dimension (L×W×H) (mm) | 440x360x160 | |
Weight (kg) | 12 | |
Environmental conditions | ||
Operating temperature | 20°C±10°C | |
Relative humidity | ≤85%RH |