GF6018 multimeter instrument calibrator
Main Functions and Features
1. The device consists of a single-phase AC voltage standard resource, current standard resource and DC voltage, current standard resource of standard resource.
2. Have calibration function, which can accord to the standard of external value, fixed standard resource output values.
3. The setting of the range for verification 1999 (9), 3999 (9) a display of the digital multimeter.
Frequency output
Sine wave output range (using AC voltage, current output provide)
(45-65) N Hz (N = 1, 2, 3, ...31);
Fineness 0.001 N Hz (N = 1, 2, 3, ...31);
The setting accuracy 5 x 10-5;
Point 1 KHz square wave frequency output, 5 KHz, 10KHz, 50KHz, 100KHz, 200KHz, 1000KHz;
Square wave amplitude 4.5 V;
Electronic resistance box output
Quantity process
0 ~ 200 Ω , 200 ~ 400 Ω , 0 ~ 2 K Ω , 2 K Ω ~ 4 K Ω , 0 ~ 20 K Ω , 20 K Ω ~ 40 K Ω , 0 ~ 200 K Ω , 200 K Ω ~ 400 K Ω , 0 ~ 2 M Ω , 2 M Ω ~ 4 M Ω ;
Fineness 0.025%;
Setting accuracy ± (0.12% RD + 0.08% FS).
Work power
Single-phase AC 220 V± 10%, Frequency 50/60 Hz
Use of the environment
Relative humidity ≤ 85% RH.
Volume and weight
440 mm x 450 mm x 160 mm, about 12 kg
Our Service
We are able to provide Complete set of scan head according to your detailed requirements, please contact me directly by skype or send inquiries to me at any time! We are always in a position to quote you the most advantageous price for higher quality merchandise. ^_^
Current/voltage
output |
AC |
DC |
Voltage range |
200 mA-1000 |
200 mA-1000 |
Current range |
20μA--20A |
20μA--20A |
Adjusting range |
0 ~ 120% |
0 ~ 120% |
Adjust fineness |
5 x 10-5 |
5 x 10-5 |
output capacity |
20 VA |
Current: 30W |
Voltage: 20W |
adjustment for accuracy |
± (0.03% RD + 0.02% FS) |
± (0.03% RD + 0.02% FS) |
stability |
≤ 0.01% / 60 s |
≤0.01% / 60 s |
waveform distortion degree |
≤0.3% |
|
output ripple |
|
≤1%. |