Customization: | Available |
---|---|
After-sales Service: | 24 Hour Service |
Warranty: | 3 Years |
Still deciding? Get samples of US$ 50000/Set
Request Sample
|
Suppliers with verified business licenses
Audited by an independent third-party inspection agency
0-600A/0-1200V;
8 inch TFT touch screen;
Accuracy 0.02% or 0.01%;
Using 24bit A/D sampling technology;
Recorder 10000 sets energy meter data;
Electrical isolation between U and I channels;
Programmable multi-plan for testing DC meter;
ISO17025 electrical metrology laboratory standard;
User friendly menu guided operation;
As high precision DC reference standard;
DC electric energy accumulation function;
Easy verification and analysis of DC meter;
Testing all kinds of DC ammeter, DC voltmeter;
Statistical U & I of maximum and minimum values;
Automatic operation without need of an external PC;
Testing all kinds of DC energy meter, DC power meter;
Electrical parameters | |
---|---|
Accuracy | 0.05%, 0.02%, 0.01% |
Power Supply | One Phase AC 180-265V, frequency 50/60Hz. |
Power consumption | 20VA |
DC Voltage Measurement | |
Range | 10mV, 100mV, 1V, 10V, 30V, 100V, 300V, 1000V (max 1200V) |
Accuracy | ±(0.012%RD+0.008%RG) |
DC Current Measurement | |
Range | 0.7mA, 3.5mA, 14mA, 35mA, 140mA, 700mA, 1A, 2A, 5A, 10A, 20A, 50A, 100A, 500A (max 600A) |
Accuracy | ±(0.012%RD+0.008%RG) |
DC Power Measurement | |
power accuracy | ±0.02%, 0.01% |
DC Power Energy Measurement Error | |
power energy | ±0.02%, 0.01% |
Power Pulse Output | |
Power Pulse Output | 20KHz |
Power Pulse Input | |
Energy pulse input | 20KHz |
Function | |
LCD Display | 8 inch 800x600 touch TFT |
Energy accumulation | Yes |
Self-calibration | Yes |
Data storage | Yes |
PC software | Optional |
Communication port | USB, RS232, 10/100M LAN |
Standard | |
Standard | IEC 62053-21,22, 23; IEC 60736; ANSI C12.20-2002; JJG-842-2017; JJG596-2012; JJG 1085-2013; JJF 68-2019; DL/T 1478-2015; DL/T 448-2016; GB/T 33708-2017 |
Safety | |
Isolation protection | IEC 61010-1:2001 |
Measurement Category | 300 V CAT III, 600 V CAT II |
Degree of protection | IP40 |
Declaration of conformity | CE & CNAS certified |
Mechanical parameters | |
Dimensions (W×H×D) (mm) | 400x440x290 |
Weight (kg) | 12 |
Environmental conditions | |
Ambient temperature | -10°C to +40°C |
Storage temperature | -20°C to +65°C |
Relative humidity | 10%-85% |
Temperature coefficient | ≤0.0005% /ºC |
Influence of external fields | ≤0.05 %/mT |
(1) RG means range, the same as below; (2) RD means the setted harmonic content, harmonic can be a single output, also multiple output. |