Portable Test-200 Electronic Measuring Instrument 200A Primary Current Injection for Test Equipment

Product Details
Customization: Available
Certification: CCC
Protection: Corrosion Resistance
Still deciding? Get samples of $ !
Request Sample
Diamond Member Since 2013

Suppliers with verified business licenses

Audited Supplier Audited Supplier

Audited by an independent third-party inspection agency

Registered Capital
5000000 RMB
Plant Area
>2000 square meters
  • Portable Test-200 Electronic Measuring Instrument 200A Primary Current Injection for Test Equipment
  • Portable Test-200 Electronic Measuring Instrument 200A Primary Current Injection for Test Equipment
  • Portable Test-200 Electronic Measuring Instrument 200A Primary Current Injection for Test Equipment
Find Similar Products

Basic Info.

Model NO.
TEST-200
Installation
Hanging
Housing
UPR
Accuracy
±0.5%
The Current Output Waveform Distortion
≤5%
Capacity
2kVA
Transport Package
Wooden Box
Specification
Phasor checker: 187×119×48mm Detector
Trademark
GFUVE
Origin
Beijing
HS Code
9107000000
Production Capacity
1000

Product Description

TEST-200200A PORTABLE PRIMARY CURRENT INJECTION TEST KIT

TEST-200 200A primary current injection test kit adopts ARM Chip to control its output process and large capacity of toroidal transformers. It is equipped with current meter; it can show primary current value, second current value, the action time and voltage value. Packing is used as aluminum alloy body with PC panel. LCD screen display single-phase protective relay test device of voltage and current output stopwatch; a full isolation, adjustable AC and DC voltage, AC direct current; an adjustable DC voltage and an adjustable AC voltage; output circuit with double overload protection, high output capacity, small size, light weight, high reliability.
Primary Injection Kit Compact Model TEST-200 is designed to operate on single phase 230/240 Volts AC 50/60 Hz. to develop smooth & continuous test current. The test set gives high current output at low cost and with inherent safety at minimum precaution. TEST-200 is mainly applied to Circuit Breaker testing in the power system, contact resistance test and relay test.
Portable Test-200 Electronic Measuring Instrument 200A Primary Current Injection for Test EquipmentPortable Test-200 Electronic Measuring Instrument 200A Primary Current Injection for Test EquipmentPortable Test-200 Electronic Measuring Instrument 200A Primary Current Injection for Test Equipment
 

FEATURES

DC Voltage output up to 300V;

AC Voltage output up to 250V;

Max current output up to 250A;

Built-in digital timer resolution 1ms;

Thermal and short-circuit protection;

Permanent current up to 200 A, 2KVA;

Built-in digital ammeter 0.2% accuracy;

Small size, light weight, easy to test on-site;

Suite for all kinds of overcurrent relay testing;

With high precision 0.0000-99999S Stopwatch;

Current injection continuous output from 0-50A;

Have one channel DC Auxiliary Output from 0 to 220V;

 

FEATURES

DC Voltage output up to 300V;

AC Voltage output up to 250V;

Max current output up to 250A;

Built-in digital timer resolution 1ms;

Thermal and short-circuit protection;

Permanent current up to 200 A, 2KVA;

Built-in digital ammeter 0.2% accuracy;

Small size, light weight, easy to test on-site;

Suite for all kinds of overcurrent relay testing;

With high precision 0.0000-99999S Stopwatch;

Current injection continuous output from 0-50A;

Have one channel DC Auxiliary Output from 0 to 220V;

 

PARAMETERS

Electrical parameters
Accuracy ±0.5%
Power Supply (AC) 220V±10%·2000VA, 50Hz/60Hz or110V±10%·2000VA, 50Hz/60Hz
Current Output (AC) 200A·5V or 100A·10V
Capacity 2KVA
The current output waveform distortion ≤5%
Dual protection overload electronic protection and software protection
Primary current
Range 0-250A
Resolution 0.1A
Accuracy <0.5% RG
AC Voltage Output
Range 0-250V AC
Resolution 0.1V
Accuracy <0.2% RG
DC Voltage Output
range 0-300V DC
Resolution 0.1V
Accuracy <0.5% RG
Current Output duration
200A range 200A  1-2minutes
100A  3-6minutes
50A  Continuous
10A range 10A  Continuous
Stopwatch
Range 0.0000-99999S
Resolution 0.1mS
Accuracy ±5 words
Max input voltage
(With contact and potential input)
DC 250V
Ammeter
Range Inside 0.000-250.0A
Outside 0.000-6.000A(AC or DC)
Accuracy <0.2%
Voltmeter
Range 0.0-600.0V(AC or DC)
Accuracy <0.2%
Measuring mode AC True RMS
DC Average value
Auxiliary Contacts
Binary input/output one, AC250V*1A or DC120V*1A
DC Auxiliary Output
Range 0-110V DC, max 0.5A; 110-220V DC, max 0.5A
Interface
Serial interface RS232
LCD 320*240 Greenlight LCD display
Resistor
Resistance 0.5Ω-2.5kΩ
Capacitor 10 μF, max voltage 450 V AC
Standard
Standard 2006/95/EC, EMC 2004/108/EC, IEC61000, IEC61010, IEC 61326
Mechanical parameters
Dimensions (W×D×H) (mm) 340×270×250
Weight(kg) 22.5
Environmental conditions
Operating Temperature -10°C to 50°C
Storage temperature -40°C to 70°C
Humidity range Relative humidity 5 ... 95 %, non-condensing

SELECTION GUIDE

Model Accuracy Primary Current Ammeter(0-6A) Timer Power Injection Current time Dimension(W×D×H) (mm) Weight Remark
TEST-200 0.50% 0-200A YES YES 2KVA 200A: 2-3min;
100A:3-6min
340×270×250 22.5KG Integrated structure
TEST-901 0.50% 0-1100A YES YES 5KVA 1000A: 2-3min;
500A:3-6min
430×260×260 25KG Integrated structure
TEST-902 0.50% 0-2200A YES YES 12KVA 2000A:2-3min;
1000A:3-6min
360×300×295 and 320×300×300 35KG+15KG Fission structure
TEST-903 0.50% 0-3300A YES YES 15KVA 3000A: 2-3min;
1500A:3-6min
380×350×360 and 340×290×420 48KG+42KG Fission structure
TEST-905 0.50% 0-5500A YES YES 30KVA 5000A:2-3min;
2500A:3-6min
430×430×445 and 550×450×600 68KG+55KG Fission structure


 

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier
Product Groups
More